Correct Answer - Option 2 : 100
Given that,
Sensitivity = 30 mV/kPa
Range = 0 kPa to 100 kPa
Full scale voltage = 30 × 10-3 × 100 = 3 V
Noise of readout circuit = 0.3 mV
Vref = 3.0003V
Number of bits = n = 10
\(Resolution = \frac{{{V_{ref}}}}{{{2^n} - 1}} = \frac{{3.0003}}{{{2^{10}} - 1}} = 2.933 \times {10^{ - 3}}V\)
= 0.00293V
Smallest readout = 0.00293V.
Resolution at input side \(= \frac{{0.00293}}{{30 \times {{10}^{ - 3}}}} = 97.66pa\)
≈ 100 pa